Purpose: This study aimed to evaluate the effects of fatigue and misfit on the reverse torque, load-bearing capacity, and failure modes of cemented and screw-retained two-implant-supported fixed dental prostheses (FDPs). Materials and methods: Eighty CAD-CAM zirconia frameworks with porcelain veneering were assigned to three groups based on engaging (E) and non-engaging (N) abutment configurations: N-N, N-E, and E-E. Misfit scenarios included a control cast and 100 µm horizontal or vertical misfits. Frameworks were mounted on implants as screw-retained or cement-retained 3-unit FDPs, forming seven test groups and one control group (n = 10). Reverse torque was measured before and after thermomechanical cycling, followed by maximum failure load (Fmax) testing and failure mode analysis. Fmax was evaluated using two-way ANOVA, and reverse torque using mixed-effects ANOVA accounting for engagement, misfit, and implant position (α = 0.05). Conclusion: The engagement configuration of implants did not significantly affect reverse torque, fracture strength, or failure modes of cemented and screw-retained restorations, although horizontal misfit reduced load-bearing capacity.