Early EGM patterns, specifically Dipper- or Riser-type fluctuations, can predict favorable R-wave amplitude increases within 6 minutes post-fixation during pacing lead implantation.
Does early EGM pattern predict favorable intraoperative R-wave amplitude increases in patients undergoing pacing lead implantation?
Early EGM patterns within 6 minutes of pacing lead fixation can predict favorable R-wave amplitude increases, potentially streamlining intraoperative lead assessment.
Absolute Event Rate: 0% vs 0%
Most intraoperative R-wave amplitude fluctuations in successful lead implantations follow Dipper- or Riser-type patterns. Favorable R-wave amplitude increases can be predicted within 6 minutes post fixation using early EGM patterns.
Oshima et al. (Wed,) reported a other. Early EGM patterns, specifically Dipper- or Riser-type fluctuations, can predict favorable R-wave amplitude increases within 6 minutes post-fixation during pacing lead implantation.