Background Point-of-care testing (POCT) using blood gas analyzers provides rapid results and can reduce diagnostic delays. Discrepancies between POCT and central laboratory (CL) measurements have been observed, particularly in patients with sodium disorders. This study evaluated the agreement between POCT and CL measurements for sodium and potassium in patients with hyponatremia, normonatremia and hypernatremia.Methods In this retrospective single-center analysis, all paired measurements of sodium and potassium performed by POCT (Radiometer ABL 90 Flex Plus) and the CL (Roche Cobas Pro ISE) within ± 2 h between January and December 2024 were included. The correlation, Bland-Altman analysis, paired tests and intraclass correlation coefficients (ICC) were used to assess agreement. Regression analysis evaluated proportional bias and ANOVA tested differences among dysnatremic and normonatremic groups.Results A total of 6404 paired sodium and 5622 paired potassium measurements were analyzed. Sodium was 140.3 mmol/L (SD 4.5) (POCT) and 137.8 mmol/L (SD 6.6) (CL), showing a mean difference of 2.4 mmol/L (SD 1.5). The correlation was significant (r = 0.583, p 4 mmol/L sodium or > 0.5 mmol/L potassium) occurred in 8.5% and 4.5% of cases, respectively.Conclusion While potassium showed a very strong correlation and good agreement, sodium demonstrated only moderate correlation and poor agreement according to ICC criteria. Systematic and clinically relevant deviations were frequent, particularly in dysnatremic and dyskalemic patients.