Purpose: To evaluate the short-term efficacy and optical performance of the ZEISS CT LUCIA 621P intraocular lens (IOL) in cataract patients using a diopter (D)-stratified approach. Patients and Methods: This retrospective study included 304 eyes of 216 patients who underwent phacoemulsification with CT LUCIA 621P implantation. Eyes were stratified into four groups by implanted IOL power: Group 1 ( 0.23). Photopic and mesopic defocus curves showed similar depth of focus across groups (p> 0.05 at most levels). Postoperative spherical aberration (SA) and coma remained low with no significant inter-group differences after generalized GEE adjustment (p> 0.17). Photopic CS was comparable across IOL power groups at most spatial frequencies. Conclusion: The CT LUCIA 621P IOL demonstrated consistent refractive predictability, VA across distances, and optical quality across a wide range of implanted powers, supporting its reliable use in contemporary cataract surgery. Keywords: CT LUCIA 621P, non-constant aspheric IOL, ZEISS optic, diopter-stratified analysis, visual performance, optical quality, cataract surgery
Jeong et al. (Fri,) studied this question.