Does mapping with microbipolar electrograms improve discrimination between superficial and intramural premature ventricular complexes in patients undergoing outflow tract PVC ablation?
81 patients (mean age 54 ± 11 years; 59.3% male) who underwent successful outflow tract premature ventricular complex (PVC) ablation using QDOT catheters.
Mapping with microbipolar electrograms (micro-EGMs) using QDOT catheters
Standard mapping with bipolar (bip-EGMs) and unipolar electrograms (uni-EGMs)
Ability of micro-EGM to discriminate between superficial and intramural PVCssurrogate
Microbipolar electrograms improve the accuracy of PVC mapping by aiding in depth estimation and local activation time annotation.
BACKGROUND Catheter ablation is an effective treatment for premature ventricular contractions (PVCs).Mapping typically relies on bipolar (bip-EGMs) and unipolar electrograms (uni-EGMs).Recently, catheters equipped with microbipolar electrodes have become available. OBJECTIVEThis study aimed to investigate the value of microbipolar EGMs (micro-EGMs) in PVCs mapping. METHODSWe performed a multicenter retrospective study including patients who underwent successful outflow tract PVC ablation using QDOT catheters.2 groups were identified based on the superficial or intramural origin of the PVC.EGMs-bip-, uni-, and micro-EGMswere assessed at the successful ablation sites.The primary outcome was the ability of micro-EGM to discriminate between superficial and intramural PVCs.The secondary outcome was signal amplitude.RESULTS Among 81 patients (mean age 54 6 11 years; 59.3% male), 51 had superficial and 30 had intramural PVCs.In superficial PVC, micro-EGM preceded both uni-EGM and bip-EGM in 90% of cases.In particular, when the earliest bip-EGM preceded the earliest micro-EGM by .10ms ( bip-micro .10ms), there was a high probability (sensitivity 0.94; specificity 0.75) of requiring ablation in both the right ventricular outflow tract and left ventricular outflow tract for successful PVC suppression.Micro-EGMs had higher initial deflection amplitude (1.5 vs 2.8 mV; P 5 .025).CONCLUSION Micro-EGMs allow more accurate PVC mapping, improving depth estimation.Micro-EGMs seem to be valuable for local activation time annotation with important implications on automated annotation algorithms within existing electroanatomic mapping systems.
Building similarity graph...
Analyzing shared references across papers
Loading...
Marco Bergonti
Nándor Szegedi
Paolo Compagnucci
Heart Rhythm O2
Semmelweis University
Antwerp University Hospital
Marche Polytechnic University
Building similarity graph...
Analyzing shared references across papers
Loading...
Bergonti et al. (Wed,) studied this question.
www.synapsesocial.com/papers/69df2a4be4eeef8a2a6af8ae — DOI: https://doi.org/10.1016/j.hroo.2026.03.007