Electrophoresis titration sensors are widely used for biomarker detection. However, traditional methods rely on a visible boundary for signal readout. Although conventional capacitively coupled contactless conductivity detection avoids indicator dependence, its single-point detection method suffers from long measurement times, large amounts of redundant data, and the inability to dynamically monitor the moving reaction boundary. To address these issues, we developed a novel contactless scanning capacitively coupled conductivity-detection method for microchip electrophoretic titration sensors. This method enables the rapid tracking and monitoring of the boundary within the microfluidic channel through dynamic scanning. The spatial distribution of conductivity during electrophoretic titration was theoretically analyzed. To evaluate the method, glucose was chosen as a model analyte. Quantitative detection was achieved over the linear range of 0.2–50 mM, with the limit of detection of 0.1 mM. The method exhibited satisfactory stability with relative standard deviation values ranging from 0.9% to 4.3% (n = 3). While the detection limit is higher than optical methods (0.02 mM), the results confirmed that the novel method offers merits, such as compact size, low cost and label-free operation. Moreover, it demonstrated strong potential for portable, quantitative analysis of target analytes across a wide range of applications.