PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
March 4, 2024Biosensors17 citationsOpen Access

Reagentless Glucose Biosensor Based on Combination of Platinum Nanostructures and Polypyrrole Layer

View Full Paper
NGNatalija GermanAPAnton PopovARAlmira Ramanavičienė

Key Points

Key points are not available for this paper at this time.

Abstract

Two types of low-cost reagentless electrochemical glucose biosensors based on graphite rod (GR) electrodes were developed. The electrodes modified with electrochemically synthesized platinum nanostructures (PtNS), 1,10-phenanthroline-5,6-dione (PD), glucose oxidase (GOx) without and with a polypyrrole (Ppy) layer—(i) GR/PtNS/PD/GOx and (ii) GR/PtNS/PD/GOx/Ppy, respectively, were prepared and tested. Glucose biosensors based on GR/PtNS/PD/GOx and GR/PtNS/PD/GOx/Ppy electrodes were characterized by the sensitivity of 10.1 and 5.31 μA/(mM cm2), linear range (LR) up to 16.5 and 39.0 mM, limit of detection (LOD) of 0.198 and 0.561 mM, good reproducibility, and storage stability. The developed glucose biosensors based on GR/PtNS/PD/GOx/Ppy electrodes showed exceptional resistance to interfering compounds and proved to be highly efficient for the determination of glucose levels in blood serum.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

German et al. (2024) studied this question.

synapsesocial.com/papers/68e75c84b6db6435876d2ff5https://doi.org/10.3390/bios14030134
Ask AI
Helpful
Bookmark
Share
View Full Paper