April 13, 2026Clinical Chemistry and Laboratory Medicine (CCLM)0 citations

Analytical and clinical performance of the Flash10 molecular point-of-care system for detection of influenza A and B viruses

View Full Paper
SZSheng ZengYLYanting LiYZYuling Zeng

Key Points

  • This research evaluates the effectiveness and efficiency of the Flash10 system in detecting influenza A and B viruses.
  • Tested throat swab specimens from 178 individuals with influenza-like symptoms.
  • Conducted parallel tests using Flash10, GeneXpert, and real-time quantitative PCR.
  • Assessed performance via precision, limit of detection, and specificity analysis.
  • Evaluated agreement between platforms using correlation analysis and Kappa statistics.
  • Retrieved turnaround time data for clinical workflow comparison.
  • Flash10 showed intra-batch precision of 0.79% to 1.41% and inter-batch precision of 2.62% to 3.05%.
  • Limit of detection for both FluA and FluB was established at 500 copies/mL with a detection rate of 100%.
  • Strong correlation with RT-PCR and GeneXpert found for both FluA and FluB, with r-values ranging from 0.87 to 0.93.
  • Perfect qualitative agreement for results was observed (Kappa=1.000).
  • Median turnaround time for Flash10 was 66.0 min, significantly shorter than RT-PCR's 1,184.0 min.

Abstract

Abstract Objectives To evaluate the analytical performance and clinical workflow efficiency of the Flash10 molecular point-of-care testing (POCT) system for the detection of influenza A (FluA) and influenza B (FluB) viruses. Methods Throat swab specimens from 178 individuals presenting with influenza-like symptoms were tested in parallel using the Flash10 system, GeneXpert, and real-time quantitative PCR (RT-PCR). Analytical performance was assessed in terms of precision, limit of detection (LOD), specificity, and contamination risk. Result agreement among platforms was evaluated using correlation analysis, Bland–Altman analysis, and Kappa statistics. Turnaround time (TAT) was retrieved from the laboratory information system to compare clinical workflow efficiency between Flash10 and conventional RT-PCR. Results The Flash10 system demonstrated good precision, with intra-batch coefficients of variation (CVs) ranging from 0.79 % to 1.41 % and inter-batch CVs ranging from 2.62 % to 3.05 % at concentrations of 500 and 1,000 copies/mL. The LOD for both FluA and FluB was 500 copies/mL, with a positive detection rate of 100 %. No cross-reactivity was observed with 10 common respiratory pathogens, and no contamination was detected during testing. Ct values obtained by Flash10 showed strong correlations with those generated by RT-PCR and GeneXpert for both FluA (r=0.87–0.93, p<0.0001) and FluB (rs=0.86–0.90, p<0.0001). Bland–Altman analysis revealed consistent agreement, despite systematic Ct differences between methods. Qualitative agreement for positive and negative results was perfect (Kappa=1.000, p<0.0001). The median TAT for Flash10 was 66.0 min, significantly shorter than that of RT-PCR (1,184.0 min, p<0.0001). Conclusions The Flash10 molecular POCT system demonstrates analytical performance and result consistency comparable to established molecular diagnostic methods for the detection of FluA and FluB viruses. Its substantially shorter turnaround time supports its potential utility for rapid molecular testing in clinical point-of-care settings.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Zeng et al. (2026) studied this question.

synapsesocial.com/papers/69dc88d83afacbeac03ea9c7https://doi.org/10.1515/cclm-2026-0093