April 16, 2026Respiratory Care1 citations

Comparison of Medication Delivery by T-Nebulizer with Inspiratory and Expiratory Reservoir

View Full Paper
FPFrank M Pisut

Key Points

  • Evaluate how inspiratory and expiratory reservoirs affect medication delivery in nebulizers.
  • Used an in-vitro model to test aerosol delivery systems
  • Conducted trials with three configurations: expiratory reservoir, no reservoir, inspiratory reservoir
  • Measured aerosolized medication amount using a negative-pressure ventilator and bacteria filters
  • Group with expiratory reservoir delivered 0.22 g of medication
  • Group without reservoir delivered 0.18 g of medication
  • Group with inspiratory reservoir delivered 0.14 g of medication
  • Delivery was significantly better with the expiratory reservoir (p < 0.05)

Abstract

Using an in-vitro model, I evaluated the effect of the presence in a nebulizer system of an inspiratory or expiratory reservoir 6 inches in length (0.15 m, internal volume about 50 mL) on the amount of aerosolized medication made available to the patient. METHOD: The test apparatus included a negative-pressure ventilator that entrained aerosol from a nebulizer into a 'patient' bacteria filter that was weighed before and after each aerosol deposition. Seven trials were run with each of three system configurations: Group 1-with an expiratory reservoir, Group 2-without a reservoir, and Group 3-with an inspiratory reservoir in place. RESULTS: Weighing the filter in Group 1 indicated that 0.22 g (± 0.04) was routed to the patient, in Group 2, 0.18 g (<0.04), and in Group 3, 0.14 g (± 0.01). CONCLUSION: A T-nebulizer with an expiratory reservoir offers the patient significantly more (p < 0.05) medication than does either a T-nebulizer with an inspiratory reservoir or one without a reservoir.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Frank M Pisut (1989) studied this question.

synapsesocial.com/papers/69e07bc12f7e8953b7cbd5d7https://doi.org/10.1177/194336548903401106
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1IMPROVING THE EFFICIENCY OF DRUG ADMINISTRATION WITH JET NEBULISERS1988 · 23 citations
  2. 2Factors influencing the size distribution of aerosols from jet nebulisers.1983 · 203 citations
  3. 3Facemask or mouthpiece for delivery of nebulized bronchodilator aerosols?1981 · 32 citations
  4. 4THE EFFECT OF DIFFERENT NEBULIZERS WITH AND WITHOUT INTERMITTENT POSITIVE PRESSURE BREATHING ON THE ABSORPTION AND METABOLISM OF SALBUTAMOL1974 · 30 citations
  5. 5Efficiency of Air Compressor-Driven Nebulizers1985 · 70 citations