End-to-end testing of an online adaptive radiotherapy workflow for liver tumors using a dynamic thoraco-abdominal anthropomorphic phantom

View Full Paper
AMA. Bakhtiari MoghaddamRARaquel Figueiredo AugustoPDPatrick Darremont

Key Points

  • This research aims to validate an online adaptive radiotherapy workflow for liver tumors using a dynamic anthropomorphic phantom.
  • Constructed a TAM-ARa phantom with realistic organ models from tissue equivalent materials.
  • Performed end-to-end tests of an adaptive IMRT workflow using various deformation scenarios.
  • Acquired imaging data from CT, CBCT, and 3T MRI scans to evaluate imaging performance.
  • The phantom's imaging characteristics matched in vivo values closely, demonstrating realistic radiological features.
  • Online adaptation restored accurate dose delivery under deformation, with deviations reduced to below 3%.
  • Independent measurements confirmed sub-millimeter accuracy and coverage of planning target volume after adaptations.

Abstract

Adaptive radiotherapy (ART) requires robust end-to-end (E2E) testing tools capable of reproducing organ deformation, physiological motion, and multimodal imaging properties. The TAM-ARa (ThoracicAnthropomorphic Phantom withMotion forAdaptiveRadiotherapy) is a dynamic anthropomorphic phantom developed to validation and quality assurance of online ART workflows. Approach:The phantom was constructed with anatomically realistic bone, lung, and abdominal organ models fabricated from tissue equivalent materials. Modular components allowed simulation of interfractional and intrafractional anatomical variations, including ventilator driven respiratory motion, abdominal deformation, and variable gastric filling. CT, CBCT, and 3T MRI scans were acquired to assess imaging performance. Two E2E tests of an online adaptive IMRT workflow for liver tumors were performed on a Varian Ethos (ETHOS, Varian, USA) system. Multiple deformation scenarios were investigated, including a static reference configuration, two abdominal and one thoracic deformation for ionization chamber (IC) dosimetry, and a static plus abdominal deformation scenarios for radiochromic film dosimetry. Main results:The TAM-ARa phantom demonstrated realistic radiological characteristics, with CT Hounsfield units and MRI relaxation times closely matching reported in vivo values. The modular design allowed reproducible and controlled abdominal deformations, while setup and dosimeter placement were completed within minutes. In E2E tests for adaptive IMRT of liver tumors, excellent agreement was achieved between measured and planned dose after the treatment plan was adapted. IC E2E measurements showed that online adaptation consistently restored accurate dose delivery under deformation, reducing deviations to below 3%. Independent film measurements confirmed sub-millimeter geometric accuracy and full planning target volume coverage after adaptation, demonstrating effective compensation of deformation-induced errors. Significance:The TAM-ARa phantom provides a versatile and reproducible platform for multimodal E2E testing and validation of online ART workflows. Employing its realistic anatomical design, modular structure, and motion capabilities, TAM-ARa was successfully applied for E2E tests to validate a workflow for adaptive IMRT of liver tumors.