March 8, 2026Methods and Applications in Fluorescence1 citationsOpen Access

Normalization is essential for accurate intraoperative perfusion assessment using near-infrared fluorescence imaging.

View Full Paper
RPRoderick Camiel PeulSKSzymon M. KielbasaFSFerran Soebrata

Key Result

Normalization to maximum signal intensity reduced measurement variability to ≤1% across different NIRF camera conditions, enhancing quantitative perfusion assessment.

Key Points

  • This research investigates whether normalization can reduce measurement variability in NIRF perfusion imaging.
  • Hypothesized that normalization of fluorescence measurements corrects measurement variability.
  • Utilized a model to simulate fluorescence perfusion patterns.
  • Conducted experiments with four NIRF camera systems under varying conditions.
  • Normalization to maximum signal intensity provided consistent perfusion values (Δ≤1%).
  • In severe malperfusion cases, normalization did not sufficiently correct variability (Δ increased).

Structured PICO

Does normalization of fluorescence measurements correct for measurement variability in quantitative near-infrared fluorescence (NIRF) perfusion imaging?

P
Population
A model capable of consistently simulating fluorescence perfusion patterns
I
Intervention
Normalization of fluorescence measurements to maximum signal intensity
O
Outcome
Consistency of perfusion parameter values across varying measurement conditions (camera type, angle, rotation, and settings)surrogate

Normalization of near-infrared fluorescence imaging measurements to maximum signal intensity effectively corrects for measurement variability, reducing the need for strict standardization.

Limitations

  • In cases of severe malperfusion where peak-intensity was not reached within the measurement period, normalization did not adequately correct for measurement variability

Abstract

Significance Near-infrared fluorescence (NIRF) imaging is increasingly used for perfusion assessment in clinical care and research settings. While subjective interpretation demonstrates clinical benefits, quantitative analysis is crucial for broader adoption and classification of perfusion patterns. However, strict standardization often conflicts with fast-paced clinical workflow, obstructing broad implementation. Aim and approach This study hypothesized that normalization of fluorescence measurements could effectively correct for measurement variability, reducing the need for strict standardization in quantitative NIRF perfusion imaging. To evaluate this, a model capable of consistently simulating fluorescence perfusion patterns was employed. Experiments were conducted in an operating room using four NIRF camera systems under varying measurement conditions. Results Normalization to maximum signal intensity provided consistent perfusion parameter values across differences in camera type, angle, rotation and settings (Δ≤1%). In cases of severe malperfusion where peak-intensity was not reached within the measurement period, normalization did not adequately correct for measurement variability (Δ increased). Conclusion While standardization remains valuable beyond parameter accuracy, appropriate normalization substantially reduces dependence on strict measurement protocols. These findings support broader clinical adoption of quantitative NIRF imaging, extending its utility beyond specialized tertiary centers and facilitating widespread integration into routine surgical care. .

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Peul et al. (2026) studied this question. Normalization to maximum signal intensity reduced measurement variability to ≤1% across different NIRF camera conditions, enhancing quantitative perfusion assessment.

synapsesocial.com/papers/69ada892bc08abd80d5bba8dhttps://doi.org/10.1088/2050-6120/ae4e7c