May 8, 2026Radiology Case Reports0 citationsOpen Access

Reverse cupping–like appearance on CT: A beam hardening–related phenomenon with phantom correlation

View Full Paper
RCRohini ChoudharySPSameer Peer

Key Result

A reverse cupping-like artifact, where the central image appears relatively hyperdense compared to the periphery, was observed in abdominal CT scans of 2 patients and reproduced in a phantom model.

Key Points

  • This report aims to describe a novel reverse cupping-like artifact in CT images caused by beam hardening and to explore its physical origins.
  • CT scans of two patients' abdomens were analyzed for beam hardening artifacts.
  • A phantom experiment was conducted to reproduce the observed artifact findings.
  • The physical mechanisms behind the reverse cupping phenomenon were investigated.
  • The reverse cupping-like artifact was identified in CT images, where the center appeared hyperdense compared to the periphery.
  • Phantom experiments confirmed the presence of the artifact under controlled conditions, paralleling the clinical findings.
  • The study elucidated the relationship between beam hardening and the appearance of the artifact.

Study Design

Type

Case Report (n=2)

Structured PICO

P
Population
2 patients undergoing abdominal CT scans and a phantom model
I
Intervention
CT scanning demonstrating a reverse cupping-like artifact
O
Outcome
Observation and physical explanation of the reverse cupping-like artifact

Identifies and explains a 'reverse cupping-like' beam-hardening artifact in CT imaging, which can help radiologists avoid misinterpretation.

Abstract

The interaction of polychromatic X-ray beam with highly attenuating material leads to beam hardening which may manifest as “cupping” artifact, where the central part of the image appears relatively hypodense as compared to the periphery, a phenomenon arising out of the monochromatic beam assumption of the reconstruction algorithms. The presence of a central hyperattenuating structure with a surrounding homogeneous medium and a sufficient peripheral-to-central distance may result in a “reverse cupping-like” variant of the beam-hardening artifact, in which the central part of the image appears relatively hyperdense compared to the peripheral part. In this report, we describe a “reverse cupping-like” artifact observed in CT scans of the abdomens of 2 patients and reproduce the findings in a phantom experiment. We attempt to unearth the physical underpinnings of this phenomenon.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Choudhary et al. (2026) conducted a case report in CT scan artifact (n=2). CT scan and phantom experiment was evaluated on Observation and reproduction of reverse cupping-like artifact. A reverse cupping-like artifact, where the central image appears relatively hyperdense compared to the periphery, was observed in abdominal CT scans of 2 patients and reproduced in a phantom model.

synapsesocial.com/papers/69fd7ddcbfa21ec5bbf0615fhttps://doi.org/10.1016/j.radcr.2026.04.003