May 21, 2026European Heart Journal - Case Reports0 citationsOpen Access

Thoracoscopic right atrial access for septal lead implantation: a novel minimally invasive technique (case report)

View Full Paper
EVEvelyne VerhulstEBElise BakelantsFHFilip Haenen

Key Result

Thoracoscopic right atrial access for septal lead implantation was successfully performed with stable parameters in a 67-year-old patient lacking conventional venous access.

Key Points

  • To introduce a novel minimally invasive technique for septal pacing lead implantation via right atrial access for patients with venous access limitations.
  • Described a novel hybrid thoracoscopic approach for right atrial access under general anesthesia.
  • Utilized three 5 mm ports for pericardial access and lead positioning using fluoroscopic and electrophysiological guidance.
  • Conducted the procedure on a 67-year-old patient with failed conventional CRT implantation due to venous stenosis.
  • Successfully implanted a pacing lead in the left bundle branch area with satisfactory electrical parameters.
  • Postoperative recovery was uneventful, and lead parameters remained stable during follow-up interrogation.
  • Demonstrated technical feasibility and procedural safety for this novel technique.

Study Design

Type

Case Report (n=1)

Structured PICO

Is thoracoscopic right atrial access feasible for septal lead implantation in a patient without conventional venous access?

P
Population
A 67-year-old patient with heart failure with reduced ejection fraction and left bundle branch block, referred for cardiac resynchronization therapy, with bilateral subclavian venous stenosis and unsuitable coronary sinus lead positioning.
I
Intervention
Hybrid thoracoscopic approach for right atrial access and implantation of a septal pacing lead targeting the left bundle branch (LBB) area.
O
Outcome
Successful lead placement with satisfactory electrical parameters and uneventful postoperative recovery.

Thoracoscopic right atrial access is a feasible minimally invasive strategy for septal lead implantation in patients lacking conventional venous access.

Limitations

  • Further experience and long-term follow-up are required to determine the reproducibility and durability of this approach.
  • Initial experience
  • Further experience and long-term follow-up required to determine reproducibility and durability

Abstract

Abstract Background Transvenous lead implantation remains the standard of cardiac pacing; however, alternative strategies are required in patients with limited or absent venous access. Video-assisted thoracoscopic surgery (VATS) has previously been used for epicardial left ventricular lead placement when transvenous approaches fail. Direct thoracoscopic access to the right atrium for endocardial septal pacing has not been previously described. We report a novel, minimally invasive technique allowing right atrial access and implantation of a septal pacing lead targeting the left bundle branch (LBB) area. Case summary A 67-year-old patient with heart failure with reduced ejection fraction and LBB block was referred for cardiac resynchronization therapy. Conventional transvenous CRT implantation was unsuccessful due to bilateral subclavian venous stenosis and unsuitable coronary sinus lead positioning with high pacing thresholds and phrenic nerve stimulation. Surgical epicardial lead placement was considered high risk because of extensive intrathoracic adhesions following previous oesophageal surgery. A hybrid thoracoscopic approach was therefore pursued. Under general anaesthesia with single-lung ventilation, three 5 mm thoracoscopic ports were inserted on the right anterior thorax. After pericardial opening, a purse-string suture was placed on the right atrium, allowing introduction of a 7–9 Fr sheath using the Seldinger technique. Under fluoroscopic and electrophysiologic guidance, a pacing lead was advanced and positioned in the LBB area. Satisfactory electrical parameters were achieved. Postoperative recovery was uneventful, and device interrogation at follow-up demonstrated stable lead parameters and effective ventricular pacing. Discussion Thoracoscopic right atrial access may represent a feasible minimally invasive strategy for septal lead implantation in patients without conventional venous access. This hybrid surgical-electrophysiological technique allows direct atrial visualization while enabling endocardial conduction system pacing. Although this initial experience demonstrates technical feasibility and procedural safety, further experience and long-term follow-up are required to determine the reproducibility and durability of this approach.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Verhulst et al. (2026) conducted a case report in Heart failure with reduced ejection fraction and LBB block (n=1). Thoracoscopic right atrial access for septal lead implantation was evaluated on Technical feasibility and procedural safety (successful lead implantation and satisfactory electrical parameters). Thoracoscopic right atrial access for septal lead implantation was successfully performed with stable parameters in a 67-year-old patient lacking conventional venous access.

synapsesocial.com/papers/6a0ea17cbe05d6e3efb60334https://doi.org/10.1093/ehjcr/ytag319