Late Dislodgement of a Leadless Pacemaker: Potential Role of Impedance Decline as an Early Warning Sign.

View Full Paper
TGToshihiko GotoMYMasashi YokoiKMKento Mori

Key Points

  • Pacing failure occurred following significant changes in pacing impedance and threshold over the initial two weeks.
  • Impedance decreased sharply within one week, indicating a potential early warning sign for dislodgement risk.
  • The observation of threshold increase at the two-week mark raises concerns about device performance post-implantation.
  • The findings suggest monitoring impedance changes could help in timely intervention and device management.

Abstract

Pacing impedance declined rapidly 1 week after implantation, followed by a significant increase in threshold at the 2-week follow-up, resulting in pacing failure.