April 17, 2026Scandinavian Journal of Trauma Resuscitation and Emergency Medicine0 citationsOpen Access

Sonorheometry to detect traumatic induced coagulopathy during initial management of trauma patients: an observational, multi-center study

GDGary DuclosPRPierre Antoine Seube RemyMWMathieu Willig

Key Points

  • The study aimed to validate the Quantra device for detecting trauma-induced coagulopathy in severe trauma patients.
  • Multicenter analysis of data from four level-1 trauma centers.
  • Simultaneous testing of Quantra and standard laboratory tests upon hospital admission.
  • Assessment of diagnostic performance using receiver operating characteristic curves.
  • Application of Youden-optimized cutoffs in an external validation cohort.
  • In the derivation cohort, AUCs for clot time predicting aPTT ratio > 1.5 was 0.91.
  • Clot stiffness had an AUC of 0.82 for predicting PTr > 1.5.
  • Fibrinogen contribution to clot stiffness had an AUC of 0.87 for predicting fibrinogen concentration < 1.5 g/L.
  • In the validation cohort, negative predictive values were high, especially for clot time at 1.00.

Abstract

Abstract Background In severe trauma 30% of the patients develop a trauma-induced coagulopathy (TIC), which is associated with higher mortality, massive transfusion and prolonged intensive care stay. Conventional coagulation assays present limitations to detect TIC, primarily due to delayed turnaround time. To enhance early management of TIC viscoelastic tests have been included into guidelines as a standard of care. The Quantra device, based on sonorheometry analysis, offers a rapid bedside assessment of coagulation but lacks validation in a trauma population. The aim of this study was to validate Quantra to detect TIC compared with standard blood coagulation test in severe trauma patients. Methods We performed a multicenter analysis of prospectively collected data (January 1, 2022–December 31, 2024) across four level-1 trauma centers. Trauma patients > 16 years underwent simultaneous Quantra and standard laboratory testing at hospital admission. Diagnostic performance for prespecified thresholds—activated partial thromboplastin time (aPTT) ratio > 1.5, prothrombin time ratio (PTr) > 1.5, fibrinogen ≤ 1.5 g/L, and platelet concentration ≤ 50, ≤ 100, and ≤ 150 G/L—was assessed by receiver operating characteristic curves in a derivation cohort, with Youden-optimized cutoffs applied to an external validation cohort to determine their specific positive and negative predictive values. Results The derivation and validation cohorts included 285 and 219 patients, respectively. In the derivation cohort, the areas under the curve (AUCs) were 0.91 0.84–0.98 for clot time to predict an aPTT ratio > 1.5 with a best cutoff of 144.5s; 0.82 0.75–0.90 for clot stiffness to predict a PTr > 1.5 (best cutoff 15.4hPa); 0.87 0.81–0.93 for fibrinogen contribution to clot stiffness to predict a fibrinogen concentration < 1.5g/L (best cutoff 1.1hPa); 0.90 0.84–0.95 for platelet contribution to clot stiffness to predict a platelets count ≤ 100G/L (best cutoff 12.4hPa). In the validation cohort, positive predictive and negative predictive values for best cutoffs were 0.67 0.50–0.80 and 1.00 0.98–1.00 for clot time, 0.50 0.39–0.61 and 0.88 0.80–0.93 for clot stiffness, 0.58 0.44–0.70 and 0.91 0.85–0.95 for fibrinogen contribution to clot stiffness, 0.13 0.24–0.99 and 0.99 0.96–1.00 for platelets contribution to clot stiffness. Conclusion Early Quantra analysis seems to provide rapid and reliable exclusion of TIC. Prospective investigations remain required to determine usefulness in TIC management.