Development of in vitro skin corrosion test using KeraSkin™

JHJeong-hyun HongJPJungeun ParkJHJee-hyun Hwang

Key Points

  • This research aims to evaluate KeraSkin™ as a reliable in vitro model for skin corrosion testing according to OECD TG 431.
  • Evaluated 30 reference chemicals using KeraSkin™ for skin corrosion testing.
  • Measured cell viability post-exposure with MTT assay after 3 or 60 minutes.
  • Applied OECD TG 431 performance standards for classification of corrosive and non-corrosive responses.
  • KeraSkin™ demonstrated 100% sensitivity and 70% specificity for predicting skin corrosion.
  • Achieved 90% overall accuracy and 76.7% sub-category accuracy in predictive capacity assessment.
  • Met OECD TG 431 performance standards for accuracy, supporting its validation as an alternative testing method.

Abstract

Skin corrosion and irritation are critical local toxicological endpoints associated with chemical exposure. OECD has approved the use of the Reconstructed Human Epidermis (RhE) model, KeraSkin™, as me-too in vitro skin irritation test of OECD Test Guideline (TG) 439. Here, we explored the applicability of KeraSkin™ as a me-too test method for in vitro skin corrosion test OECD TG 431, to expand its validated use beyond skin irritation test. Thirty reference chemicals, including Category 1A, 1B/1C, and non-corrosive substances as described in OECD TG 431 performance standards (PS), were tested. Test chemicals were applied for 3 or 60 min, and cell viability (CV) was measured using MTT assay. Classification criteria were based on viability thresholds, wherein <15% at 3 min indicates Category 1A corrosives, ≥15% at 3 min and <15% at 60 min, Category 1B/1C, and ≥50% at 3 min and ≥15% at 60 min indicates non-corrosives. The KeraSkin™ skin corrosion test demonstrated reliable predictive capacity with 100% sensitivity, 70% specificity, 90% accuracy (corrosive/non-corrosive), and 76.7% sub-category accuracy in the predictive capacity assessment using 30 chemicals, which satisfies the accuracy criteria of OECD TG 431 PS. Collectively, we demonstrated that KeraSkin™ can be a reliable in vitro model for skin corrosion test, expanding the availability of validated non-animal methods for global chemical safety assessment. Further full validation test is underway for the approval of the KeraSkin™ skin corrosion test for a me-too test method of OECD TG 431.