January 22, 2026JACOW0 citationsOpen Access

Fibre-Coupled Oscillator-Driven Electro-Optic Spectral Interferometry as a Longitudinal Bunch Profile Monitor

View Full Paper
CPC. PakuzaDGD. M. GrahamDWDavid Walsh

Key Points

  • The aim is to develop a non-destructive technique to accurately monitor the longitudinal charge profile of relativistic bunches.
  • Utilized a new technique called electro-optic spectral interferometry (EOSI).
  • Implemented measurements on the CLEAR accelerator at CERN.
  • Used a nJ oscillator with optical fibre-coupled transport.
  • Achieved successful measurements of ultrashort bunch lengths and arrival times.
  • Demonstrated low power and turn-key operation.
  • Highlighted the flexibility of integrating EOSI systems.

Abstract

Electro-optic (EO) diagnostics offer non-destructive methods to resolve the longitudinal charge profile of highly relativistic bunches without the need for complex calibrations or ambiguous phase recovery techniques. The most common technique, EO spectral decoding (EOSD), is favoured for its simplicity, reliability, and straightforward output interpretation, however its resolution is constrained to the geometric mean of the transform-limited and chirped probe laser durations. We have introduced a new technique utilising spectral interferometry (EOSI), which overcomes this limitation by adding a single optical element to an EOSD setup, with successful measurements shown previously on the CLARA accelerator at Daresbury Laboratory. To further explore this technique, we have performed EOSI on the CLEAR accelerator at CERN (160-200 MeV, 15-150 pC and 0.2-1.5 ps RMS) utilising a nJ oscillator and optical fibre-coupled transport. Our results highlight the potential for low power, turn-key operation and flexible integration of EOSI systems for single-shot ultrashort bunch length and arrival time monitoring.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Pakuza et al. (2026) studied this question.

synapsesocial.com/papers/6971bdad642b1836717e2608https://doi.org/10.18429/jacow-ibic2025-wepco40
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Compact electro-optical bunch length detector: from an expert device to an operator tool2026
  2. 2Compact electro-optical bunch length detector: from an expert device to an operator tool2025
  3. 3An electro-optical bunch profile monitor for FCC-ee2025
  4. 4Electro-optically sampled broadband dual-comb spectroscopy across mid-IR to terahertz (6.6–200 µm) spectral range2024
  5. 5Analysis of a phase modulated interferometric Electro-Optic BPM at the CERN SPS2026