Unipolar electrogram waveforms during perforation in left bundle branch area pacing

View Full Paper
HTH TolppanenVVValérian ValitonSSS Stempfel

Key Result

Unipolar electrograms during left bundle branch area pacing perforation exhibited wide and deep Q/S waves in 86% of cases, compared to 15% in the final accepted lead position.

Key Points

  • This research aimed to explore unipolar electrogram waveforms during perforation events in left bundle branch area pacing.
  • Analyzed a prospective database of patients undergoing LBBAP implantation with timing of perforation events recorded.
  • Defined micro perforation as COI <5 mV and impedance drop >200 Ohms; macro perforation included loss of capture.
  • Recorded and compared unipolar unfiltered electrogram waveforms during perforation and at final lead position.
  • Identified three distinct EGM patterns during perforation: Q(R), RS, and R.
  • Most cases (86%) exhibited wide and deep Q/S waves, while temporary pacing cases had RS or R patterns.
  • Fifteen percent of patients had macro perforations, with notable wave characteristics differentiating them from micro perforations.

Study Design

Type

Observational (n=66)

Structured PICO

Do unipolar unfiltered EGM waveforms differ during perforation compared to the final accepted lead position in patients undergoing left bundle branch area pacing?

P
Population
66 patients (mean age 80 years, 38% female) undergoing left bundle branch area pacing complicated by perforation with available timestamped electrograms.
E
Exposure
Evaluation of unfiltered (0.5 – 500 Hz) unipolar sensed tip EGM waveforms during perforation
C
Comparator
Evaluation of unfiltered unipolar sensed tip EGM waveforms in the final accepted lead position in the same patient
O
Outcome
Waveforms of unipolar unfiltered EGM (negative and positive deflections 'Q/S' and 'R' waveforms, and current of injury [COI] amplitudes)surrogate

A deep and wide negative deflection with low current of injury amplitude on unipolar EGM can help identify lead perforation during left bundle branch area pacing.

Main Result

Absolute Event Rate: 86% vs 15%

Abstract

Abstract Introduction Perforation is common during left bundle branch area pacing (LBBAP) implantation, as the conduction fibers lie just beneath the left ventricular endocardium. Whereas overt (macro) perforation is usually straightforward to diagnose due to loss of capture, recognizing micro perforation, in which only the tip of the lead screw perforates, is much more challenging as thresholds are usually preserved. The main parameters to evaluate perforation are pacing impedance and current of injury (COI) amplitude from the unipolar unfiltered electrogram (EGM). Besides COI amplitude, there are few data on distinct EGM waveforms which are related to perforation. Purpose Our aim was to investigate waveforms of unipolar unfiltered EGM during perforation and compare them to those of the final accepted lead position. Methods All patients with LBBAP implantation in whom perforation was indicated in our prospective database and which was annotated by a timestamp in the recording system were included in the analysis. Micro perforation was suspected in case of COI 5 mV together with a drop in impedance by 200 Ohms. Macro perforation was defined if in addition there was loss of capture at 5V/0.5ms. Unfiltered (0.5 – 500 Hz) unipolar sensed tip EGM waveform during perforation was compared to that in the final lead position in the same patient. The negative and positive deflections ("Q/S" and "R" waveforms) and COI amplitudes were measured by digital calipers of the recording system. Results There were 120/622 cases with at least one perforation, of which 66 (mean age 80 ± 11 years, 62% men) had a time stamp which enabled measurement of the EGM at the time of perforation and at the final position. We recognized 3 distinct patterns of EGMs during perforation: Q(R), RS and R. Most cases (86%) had wide and deep Q/S waves (Table). No Q waves were visible if the sensed beats originated from temporary pacing (n=11, 17%); in those cases, the pattern was RS (n=6) or R (n=5). In the final accepted position, the most common pattern was R with large COI; although 15% patients had a Q wave it was markedly shorter in duration and of lower amplitude than in the perforated beats (Figure). Ten (15%) patients had macro perforations; the remaining 56 (85%) patients had micro perforations. In those with macro perforation, all cases without temporary pacing showed a wide and deep Q wave (67ms, IQR 66-70ms; 11.9mV, IQR 11.2-12.4mV). In those with micro perforation, 31 had a Q(R) pattern (Q-wave duration 56ms, IQR 50-68mV; 6.0mV, IQR 4.4-10.6mV), 14 had an RS pattern (four with temporary pacing) and eight had an R pattern (four with temporary pacing). The cases with RS or R pattern had a small R (2.8mV, IQR 1.5-3.8mV) and low COI (3.4mV, IQR 1.7-4.8mV). Conclusions The unipolar EGM from lead tip showed very distinct patterns during perforations compared to the final position. A deep and wide negative deflection, in addition to low COI amplitude, is indicative of perforation.Table Figure

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Tolppanen et al. (2025) conducted an observational in Left bundle branch area pacing (LBBAP) implantation (n=66). Perforation during left bundle branch area pacing vs. Final accepted lead position was evaluated on Presence of wide and deep Q/S waves on unipolar unfiltered electrogram. Unipolar electrograms during left bundle branch area pacing perforation exhibited wide and deep Q/S waves in 86% of cases, compared to 15% in the final accepted lead position.

synapsesocial.com/papers/698585bd8f7c464f230094c8https://doi.org/10.1093/eurheartj/ehaf784.647
Ask AI
Helpful
Bookmark
Share
View Full Paper