February 6, 2026European Heart Journal0 citations

Evaluating right ventricular septal implantation ratio of Micra TPS using post-procedural CT and predictive factors

View Full Paper
KYK YanoMSM ShinoharaKSK. Shibata

Key Result

Maintaining adequate distance from the RV free wall and forming an α-loop shape independently predicted successful Micra TPS RV septal implantation (OR 20.1; 95% CI 4.68-86.6; p<0.001).

Key Points

  • The study aims to identify the ratio of right ventricular septal implantation of the Micra device and factors that influence successful implantation.
  • Enrolled 110 patients who underwent Micra TPS implantation.
  • Conducted post-procedural CT scans to evaluate the implantation site.
  • Defined successful RV septal implantation versus non-septal regions.
  • Assessed delivery catheter positioning and the formation of an α-loop shape.
  • RV septal implantation observed in 74 patients (67.3%).
  • 36 patients (32.7%) had non-septal implantation positioning.
  • Successful RV septal implantation maintained a distance greater than half the Micra size from the RV walls.
  • The formation of the α-loop shape was highlighted as a strong predictor with an odds ratio of 20.1.

Study Design

Type

Cohort (n=110)

Structured PICO

Does maintaining adequate distance from the RV free wall and forming an α-loop shape during delivery improve the rate of successful right ventricular septal implantation of the Micra TPS?

P
Population
110 consecutive patients who underwent Micra Transcatheter Pacing System (TPS) implantation
I
Intervention
Delivery catheter positioning maintaining an adequate distance from the right ventricular (RV) free wall and forming an α-loop shape during Micra TPS implantation
C
Comparator
Delivery catheter positioning without adequate distance from the RV free wall or without forming an α-loop shape
O
Outcome
Successful right ventricular (RV) septal implantation confirmed by post-procedural CT scan (defined as not being positioned in the non-septal region such as RV apex, free wall, or inferior wall)surrogate

Maintaining adequate distance from the RV free wall and forming an α-loop shape with the delivery catheter significantly increases the likelihood of successful right ventricular septal implantation of leadless pacemakers.

Main Result

Effect estimate: OR 20.1 (95% CI 4.68-86.6)

p-value: p=<0.001

Abstract

Abstract Introduction A septal implantation of the Micra Transcatheter Pacing System (TPS) is recommended to avoid myocardial injury. However, there are no reports to evaluate the actual Micra device implantation site using post-procedural Computed tomography (CT) scans. This study aimed to identify the right ventricular (RV) septal implantation ratio and factors regarding successful RV septal implantation of Micra device. Methods A cohort of 110 consecutive patients who underwent Micra TPS implantation were enrolled. CT scans after procedure were performed to evaluate the actual implantation site of the Micra device. RV septal implantation was defined as not being positioned in the non-septal region such as RV apex, the free wall, or the inferior wall of the RV. Delivery catheter positioning, specifically maintaining an adequate distance from the RV free wall and/or forming an α-loop shape during the procedure were evaluated (Figure A and B). Results Post-procedural CT scans revealed RV septal implantation in 74 patients (67.3%) and non-septal positioning in 36 patients (32.7%). Compared to cases with non-RV septal implantation, in those with RV septal implantation, the distance between the Micra device and the RV walls was maintained more than half the size of the Micra device in the right anterior oblique view, as well as formation of the α-loop shape in the left anterior oblique view. Multivariate analysis revealed that "adequate distance and the α-loop shape" was the independent predictor of successful RV septal implantation (OR:20.1 4.68-86.6, p 0.001, Figure C). Conclusions Successful ratio of RV septal implantation was slightly lower than expectation. Maintaining a distance between the tip of Micra and RV free wall and the forming an α-loop shape of the delivery catheter are important factors for successful implantation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yano et al. (2025) conducted a cohort in Patients undergoing Micra Transcatheter Pacing System (TPS) implantation (n=110). Delivery catheter positioning (adequate distance from RV free wall and α-loop shape) vs. Absence of adequate distance and α-loop shape was evaluated on Successful right ventricular (RV) septal implantation (OR 20.1, 95% CI 4.68-86.6, p=<0.001). Maintaining adequate distance from the RV free wall and forming an α-loop shape independently predicted successful Micra TPS RV septal implantation (OR 20.1; 95% CI 4.68-86.6; p<0.001).