February 8, 2026ACS Sensors0 citations

Clinical Evaluation of Microneedle Biosensors for Continuous Lactate Monitoring in Critically Ill Patients

View Full Paper
ODOmeed DjassemiACAn-Yi ChangWMW. Cameron McGuire

Key Points

  • This research aims to evaluate a microneedle biosensor for continuous lactate monitoring in critically ill patients to improve patient management.
  • Developed a microneedle-based electrochemical biosensor for lactate monitoring.
  • Conducted a pilot study with 21 participants in varied clinical settings.
  • Used two enzyme-based sensors placed on forearm and thigh for continuous tracking of lactate.
  • Measured agreement with traditional blood lactate assays using statistical analysis.
  • Demonstrated strong agreement between biosensor and blood lactate assays (r = 0.94).
  • Achieved high diagnostic accuracy for hyperlactatemia with an area under the curve of 0.95.
  • Showed minimal bias (-0.028 mmol/L) over a broad measurement range (0.7-22.9 mmol/L).
  • No significant differences between interstitial fluid and blood lactate levels were found.

Abstract

Continuous lactate monitoring is critical for early detection and management of sepsis, shock, and metabolic stress, yet current serum assays remain invasive, intermittent, and resource-intensive. We present a clinical evaluation of a minimally invasive microneedle-based electrochemical biosensor for real-time interstitial fluid (ISF) lactate monitoring. The microneedle biosensor features a platinum working electrode modified with a lactate oxidase reagent layer and a polyvinyl chloride anti-fouling membrane for H2O2-mediated amperometry, toward highly selective and stable ISF lactate detection. In a pilot study of twenty-one participants across an intensive care unit, emergency department, cardiopulmonary exercise testing, and controlled laboratory settings, two enzyme-based microneedle sensors placed on the forearm and thigh continuously tracked lactate for 4 h. Sensor performance demonstrated strong agreement with blood lactate assays (r = 0.94), high diagnostic accuracy for hyperlactatemia (>4 mmol/L; receiver operating characteristic analysis, area under the curve = 0.95), and minimal bias (-0.028 mmol/L) over a wide dynamic range (0.7-22.9 mmol/L) with high selectivity against interferents. No significant ISF-blood differences (p > 0.05) or adverse events were observed. These findings establish microneedle biosensors as a promising platform for precision medicine, with considerable potential to transform sepsis care, guide resuscitation, and improve assessment of exertional dyspnea.