February 11, 20260 citationsOpen Access

Perfusion-Dependent Melanin Bias in Pulse Oximetry and ICU Mortality Across 209 U.S. Hospitals: A Multicenter Retrospective Analysis of 52 Million Readings

View Full Paper
MGMatthew Gehring

Key Points

  • The study investigates the correlation between pulse oximetry bias and ICU mortality across diverse populations, specifically examining the effects of melanin and perfusion.
  • Analyzed 52.4 million SpO2/SaO2 readings from three data sources across 209 hospitals.
  • Defined bias as the difference between pulse oximetry (SpO2) and arterial blood gas (SaO2) readings.
  • Used multivariable logistic regression to assess mortality risk associated with hidden hypoxemia after adjusting for demographic factors and illness severity.
  • Conducted sensitivity analyses, including landmark restriction and lactate stratification.
  • Bias increased significantly under low perfusion conditions, notably in patients with high melanin.
  • Among hidden hypoxemia cases, ICU mortality was approximately twice as high compared to non-hidden hypoxemia cases (adjusted OR 1.86).
  • The identified bias in African American patients was 62% higher than expected from previous academic studies.
  • Hidden hypoxemia events occurred consistently throughout ICU stays, not just before death, with many occurring well in advance.

Abstract

Background: Pulse oximeters are typically validated on cohorts of 200–500 subjects under controlled conditions. Whether these cohorts capture the demographic heterogeneity of national clinical practice and whether measurement error is associated with patient outcomes has not been established at scale. Methods: We analyzed paired SpO2/SaO2 readings from three independent sources spanning 209 U.S. hospitals: MIMIC-IV (1 hospital; 12,934 ICU stays), eICU-CRD (208 hospitals; 55,178 stays), and the Open Oximetry Repository (PhysioNet; 52.4 million readings with continuous melanin and perfusion indices). Bias was defined as SpO2 − SaO2. Hidden hypoxemia (SpO2 ≥ 94% with SaO2 48 hours before death (median first event: 15.3 hours; mean time to death: 151 hours), and the association persisted in landmark analysis (first 48 hours only), in patients with normal lactate (adjusted OR 1.87, 95% CI 1.61–2.16), and when both restrictions were applied simultaneously (16.5% vs. 11.1%). Waveform analysis (n=125) showed no fixed racial difference in baseline PPG AC/DC ratio (Black: 0.299, White: 0.273), suggesting the signal deficit is conditional on perfusion state. Full extraction (n=1,545) is in progress. Conclusions: In this multicenter retrospective analysis, national pulse oximetry variance exceeded published benchmarks and was associated with approximately doubled ICU mortality, replicated across 209 U.S. hospitals. Hidden hypoxemia was not a pre-terminal artifact: events occurred throughout the ICU stay at a constant rate, and mortality associations persisted in landmark and lactate-stratified analyses. These findings suggest that current regulatory validation standards may underestimate the real-world prevalence of demographic bias in pulse oximetry, and that perfusion-dependent mechanisms may offer a target for algorithmic correction.