From pH to glucose: Implementing a simple point‐of‐care method for platelet quality control

View Full Paper
BVBelén del Valle VeraMVMaría‐Jesús VayáHSHéctor Sarmiento

Key Result

A point-of-care glucometer provided reliable glucose measurements in PAS-E platelet concentrates, comparable to a standard autoanalyser on day 6 (mean difference 0.4 ± 2.6 mg/dL, p=0.415).

Key Points

  • The study aimed to evaluate the feasibility of a point-of-care glucometer for measuring glucose in stored platelet concentrates.
  • Analyzed thirty platelet concentrates stored in platelet additive solutions on days 2 and 6.
  • Used a point-of-care glucometer and a reference laboratory autoanalyser for comparison.
  • Conducted statistical testing, categorical analysis, and correlation analysis.
  • Glucose levels from both methods were comparable on day 6.
  • Mean difference was 2.2 ± 5.5 mg/dL on day 2 and 0.4 ± 2.6 mg/dL on day 6.
  • Strong linear association observed with r = 0.988 and p < 0.001.

Structured PICO

Does a point-of-care glucometer provide reliable glucose measurements in PAS-E stored platelet concentrates compared to a standard laboratory autoanalyser?

P
Population
30 platelet concentrates (PCs) suspended in platelet additive solution (PAS-E)
I
Intervention
Glucose measurement using a point-of-care (POC) glucometer (Accu-Chek Instant®, Roche)
C
Comparator
Glucose measurement using a reference laboratory autoanalyser (AU5800, Beckman Coulter®)
O
Outcome
Comparability of glucose measurements between the POC glucometer and the reference autoanalyser on days 2 and 6 of storagesurrogate

A point-of-care glucometer provides reliable glucose measurements in PAS-E platelet concentrates, offering a simple and cost-effective method to implement new EDQM quality control recommendations.

Main Result

Effect estimate: Mean difference 0.4 mg/dL

p-value: p=0.415

Abstract

Abstract Background and Objectives The 21st edition of the European Directorate for the Quality of Medicines p < 0.001), correlation was not interpreted as evidence of analytical agreement. Classification of samples below the clinically relevant threshold of 10 mg/dL (≈0.56 mmol/L) did not differ significantly between methods. Conclusion A POC provides reliable glucose measurements in PAS‐E PCs, comparable to a standard autoanalyser. This simple, rapid and cost‐effective approach may facilitate implementation of the new EDQM recommendation and improve QC efficiency in BE.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Vera et al. (2026) studied Platelet concentrates stored in platelet additive solutions (n=30). Point-of-care (POC) glucometer (Accu-Chek Instant) vs. Reference laboratory autoanalyser (AU5800) was evaluated on Glucose measurement comparison on day 6 (Mean difference 0.4 mg/dL, p=0.415). A point-of-care glucometer provided reliable glucose measurements in PAS-E platelet concentrates, comparable to a standard autoanalyser on day 6 (mean difference 0.4 ± 2.6 mg/dL, p=0.415).

synapsesocial.com/papers/698d6de45be6419ac0d5318chttps://doi.org/10.1111/vox.70211