Noninvasive Measurement of Cerebrospinal Fluid Flow in Shunted Hydrocephalus: Protocol for Scanner Calibration and Multisite Data Collection

View Full Paper
JTJason M. ToliaoMBMatthew T BorzagePCPradip P. Chaudhari

Key Points

  • This study aims to establish a standardized protocol for measuring cerebrospinal fluid flow in shunted hydrocephalus using phase-contrast MRI.
  • Construct a phantom model with a human shunt catheter for testing.
  • Develop calibration curves for known flow rates versus PC-MRI measurements.
  • Assess MR scanner accuracy using linear regression analysis.
  • Conduct calibration on 8 different MR scanners before multisite data collection.
  • The phantom model for shunt flow was created by November 2024.
  • Calibration began in April 2025 and ended in December 2025.
  • Multi-institutional clinical data collection is set to start in July 2026 and last through January 2027.
  • Results and analyses are due by April 2027.

Abstract

Abstract Background Ventricular shunts divert cerebrospinal fluid (CSF) in patients with hydrocephalus, which can be lifesaving. Untreated shunt failure may lead to increased intracranial pressure and neurological injury. The process of diagnosing shunt malfunction can be complex, and historically, there has not been a simple method for noninvasive and quantitative measurement of CSF flow through shunts. The demonstration of successful clinical application of phase-contrast magnetic resonance imaging (PC-MRI) to noninvasively quantify shunt flow is relatively new and will benefit from standardization across varying types of magnetic resonance (MR) hardware to facilitate implementation at multiple medical sites. PC-MRI CSF flow measurement through ventricular shunts has not yet been compared across different types of MR hardware (ie, differing field strengths, radiofrequency coils, slew rates, gradient strengths, models, and manufacturers). Objective This study describes a protocol aimed to standardize and optimize PC-MRI shunt flow measurements for widespread use on any MR scanner. Methods To study shunt flow in a manner that would translate effectively to the patient care setting, a phantom model incorporating a human shunt catheter can be constructed to obtain data over a typical physiological range of CSF flow rates. Calibration curves are used to model data comparing known flow rates to flow measured via PC-MRI. The accuracy of each MR scanner is assessed using linear regression. This protocol will be repeated on 8 MR scanners before multisite data collection. Results The shunt flow phantom was constructed in November 2024. Single-site phantom calibration began in April 2025 and concluded in December 2025. Registration of clinical data collection sites will take place between November 2025 and July 2026. Approval from the Children’s Hospital of Los Angeles Institutional Review Board was obtained in January 2023 (before scanner calibration), and single-site data collection is ongoing. Multi-institutional clinical data collection will begin in July 2026 and continue until January 2027. Results and statistical analyses are expected by April 2027. Conclusions This study protocol provides a methodology to test and implement PC-MRI on any MR scanner using a phantom model that (1) represents real flow and catheter conditions, (2) provides a repeatable means for data collection, (3) is easy to assemble, (4) minimizes artifacts, and (5) is transportable. We also describe analytic methods for interscanner calibration across a range of hardware parameters and provide a framework for multisite data collection.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Toliao et al. (2026) studied this question.

synapsesocial.com/papers/699a9e9f482488d673cd4c90https://doi.org/10.2196/85918
Ask AI
Helpful
Bookmark
Share
View Full Paper