Both 15 kGy and 25 kGy significantly altered surface properties, but 25 kGy induced greater Ti ion release and micro-damage. Within the study limitations, 15 kGy is recommended as the preferred sterilization dose, as it maintains sterility while minimizing mechanical degradation and excessive Ti ion release as compared to 25 kGy.