Fracture of the polyethylene tibial post is a rare complication of posterior-stabilized total knee arthroplasty and is frequently missed on conventional radiography due to the radiolucent nature of polyethylene. We report a case of post-traumatic instability in which computed tomography demonstrated a displaced polyethylene fragment with characteristic low attenuation values. The diagnosis was confirmed intraoperatively, and isolated polyethylene exchange resulted in a good clinical outcome. Teaching point: CT attenuation measurements can aid in identifying displaced polyethylene fragments and facilitate the diagnosis of tibial post fracture in posterior-stabilized knee arthroplasty.
Building similarity graph...
Analyzing shared references across papers
Loading...
Cotza et al. (Thu,) studied this question.
www.synapsesocial.com/papers/69a765bfbadf0bb9e87da459 — DOI: https://doi.org/10.5334/jbsr.4202
Alexander Cotza
Filip Robijns
Jan Vandevenne
Journal of the Belgian Society of Radiology
Hasselt University
Ziekenhuis Oost-Limburg
Building similarity graph...