March 6, 2026ASAIO Journal1 citations

Venoarterial Extracorporeal Membrane Oxygenation Drainage Cannula Choice Impacts Flow Dynamics and Thrombosis Risk: A Computational Analysis

View Full Paper
AWAvishka WickramarachchiDCD. James CooperNFNavid Freidoonimehr

Key Points

  • Evaluate how cannula size, design, and flow rate influence flow dynamics and thrombosis risk in venoarterial ECMO.
  • Modeled four drainage cannulas in patient-specific vasculatures using computational fluid dynamics (CFD).
  • Simulated drainage flow rates of 2, 3, 4, and 5 L/min across 48 simulations.
  • Implemented a platelet activation model to assess thrombosis risk.
  • Measured prothrombotic flow markers and activated platelet distributions.
  • Higher flow rates worsened washout rates and increased blood residence times.
  • The 21 Fr Bio-Medicus cannula showed greater platelet activation than the 21 Fr Maquet model.
  • Patient geometries affected blood residence times and washout rates, but did not alter flow inside the cannulas.

Abstract

The drainage cannula in a venoarterial extracorporeal membrane oxygenation (VA ECMO) circuit transfers venous blood from the patient to a pump. However, drainage cannulas are prone to thrombosis, leading to significant morbidity in patients. Therefore, this study aimed to evaluate the effects of cannula size, design, and flow rate on flow dynamics and risk of flow-induced thrombosis in different patient geometries using computational fluid dynamics (CFD). Four cannulas (21 Fr and 23 Fr Bio-Medicus and Maquet models) were modeled in three patient-specific venous vasculatures. These were simulated at 2, 3, 4, and 5 L/min drainage flow rates (N = 48 simulations). To assess potential thrombosis risk, a platelet activation model was implemented along with measurement of prothrombotic flow markers. Increasing drainage flow rates worsened washout rates and increased blood residence times and platelet activation rates. The 21 Fr Bio-Medicus cannula had greater activated platelet distributions and higher tip velocities than the 21 Fr Maquet model. Lastly, patient geometry altered blood residence times, washout rates, and activated platelet distributions, but not flow within the cannulas themselves. These findings will be beneficial for clinicians in their cannula selection and cannulation strategy for VA ECMO, as well as engineers in the development of future drainage cannulas.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Wickramarachchi et al. (2026) studied this question.

synapsesocial.com/papers/69aa70c8531e4c4a9ff5ad5dhttps://doi.org/10.1097/mat.0000000000002687
Ask AI
Helpful
Bookmark
Share
View Full Paper