Monte Carlo simulations demonstrate better microcalcification detection with photon-counting detectors than energy-integrating detectors.
Microcalcification (μCalc) detection plays an important role in breast cancer screening. Electronic noise in energy-integrating detectors (EID) is the major challenge for this task in current breast cone beam CT (bCBCT) due to the tight dose constraint for breast imaging. bCBCT with a photon counting detector (PCD) can potentially offer a higher spatial resolution and lower noise. This study performed a direct comparison of bCBCTs with the two detector types via GPU-based Monte Carlo (MC) simulation. Approach. We employed VICTRE toolkit to generate a realistic breast phantom with a 0.25 mm³ voxel size, 80% fat fraction and 14 cm diameter. We considered a bCBCT system with a 60 kV x-ray source filtered with 0.3 mm Cu and detector response functions for PCD and EID. 360 projections were simulated with a total number of 3.15×10¹² photons, corresponding to ~4 mGy mean glandular dose, comparable to a two-view mammography. We modified our GPU-based MC simulation code to incorporate analytical descriptions of μCalcs of spherical shapes with diameters ranging from 0.1 to 0.4 mm, in 0.1 mm increments, into the voxelized phantom. A nichrome wire with 0.07 mm diameter was simulated to calculate the modulation transfer functions. bCBCT images were reconstructed with the Feldkamp-Davis-Kress algorithm, and image quality and μCalc detection performance were evaluated. Main results. EID-bCBCT had more profound image noise due to electronic noise. The image intensity standard deviations estimated within a region of interest were 0.055~cm⁻¹ for EID-bCBCT and 0.038~cm⁻¹ for PCD-bCBCT, respectively. μCalcs and breast anatomy such as ligaments were more visible in the PCD-bCBCT images. The 10% MTF cutoffs were 5.5 and 9.5 lp/mm for EID-bCBCT and PCD-bCBCT, respectively. Contrast-to-noise ratio ranged in 1.20-9.13 for EID-bCBCT and 3.07-14.74 for PCD-bCBCT, depending on μCalc sizes. Significance. We compared EID- and PCD-based bCBCT for μCalc detection using GPU-based MC simulations in a clinically realistic setting. Our results demonstrate a potential advantage of PCD-bCBCT for this detection task.
No takes yet. Share an insight, caveat, or question.
Ezzati et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: