DOSESTAT-QC® is a stand-alone automated quality control (QC) system used for daily quality assurance of X-ray equipment in Jönköping Region, Sweden. The software has been implemented for all mammography systems and interventional systems in the region. One of the image analysis included in the DOSESTAT-QC® software is performed in homogenous images and focuses on the standard deviations in mean pixel value (MV) and signal-to-noise ratio (SNR) in the images. Initially, the analyses were performed in 1 cm2 regions of interest (ROIs) and the obtained values in each ROI were compared to the corresponding values for the entire image. While MV remained relatively stable over time, fluctuations in SNR together with imprecise localization of pixel errors, especially in the automatic exposure control (AEC) area, highlighted limitations. In this paper, an improved method for image evaluation is presented, which enables precise SNR baseline settings and clear visualization of deviations and dead pixels. Additionally, the adaption and clinical implementation of DOSESTAT-QC® to conventional X-ray systems in the region are described.
Building similarity graph...
Analyzing shared references across papers
Loading...
Tanny Visanuyanont
Tomas Moberg
Emanuel Hillberg
Radiation Protection Dosimetry
Uppsala University
University of Gothenburg
Sahlgrenska University Hospital
Building similarity graph...
Analyzing shared references across papers
Loading...
Visanuyanont et al. (Tue,) studied this question.
www.synapsesocial.com/papers/69b6069b83145bc643d1c96e — DOI: https://doi.org/10.1093/rpd/ncaf194