PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
April 16, 2026Nanoscale Horizons0 citations

Breathable, biocompatible, long-term stability Ti 3 C 2 T x bioelectrodes for real-time monitoring of electrophysiological signals

JLJingxiao LiuZSZhaoyang SuJXJingYue Xuan

Key Points

  • The research aims to develop a bioelectrode that combines high breathability with effective monitoring of electrophysiological signals.
  • Developed Ti3C2Tx bioelectrodes with NaAsc passivation
  • Measured gas permeability and impedance at skin-electrode interface
  • Compared performance with traditional Ag/AgCl gel electrodes
  • Assessed signal-to-noise ratios for electromyography and electrocardiography signals
  • Achieved gas permeability of 1408.89 g m-2 d-1
  • Reduced skin-electrode interface impedance to 1.83 × 10^5 Ω at 10 Hz
  • Obtained a signal-to-noise ratio of 19.4 dB for electromyography signals
  • Showed superior T/R wave intensity ratio of 0.4079 for electrocardiography compared to traditional electrodes

Abstract

Electrophysiological signals, generated from the directional transport of ions across cellular membranes, encode vital physiological information relevant to human health status. Electrophysiological signal monitoring serves as a critical approach for deciphering fundamental physiological processes and enabling precise disease diagnostics. Bioelectrodes are indispensable components for the acquisition of high-fidelity electrophysiological signals. However, sweat inevitably accumulates at the skin-electrode interface during continuous monitoring, which may irritate the skin or cause the electrodes to detach from the skin, affecting the acquisition of high-quality electrophysiological signals. A key challenge in ensuring the long-term utility of bioelectrodes lies in enhancing their breathability without impeding electrophysiological signal transmission. Therefore, we realized Ti3C2Tx edge passivation by NaAsc and the preservation time at room temperature was increased to 30 days. On this basis, an antioxidant Ti3C2Tx bioelectrode with high air permeability and biocompatibility was proposed. The bioelectrode has an excellent gas permeability of 1408.89 g m-2 d-1 and a low skin-electrode interface impedance of 1.83 × 105 Ω at 10 Hz. Moreover, the signal-to-noise ratio of the electromyography signal is 19.4 dB and the T/R wave intensity ratio of the electrocardiography signal is 0.4079, which is superior to that of the Ag/AgCl gel electrode. The bioelectrode with these optimized attributes enables stable, long-term monitoring of multiple electrophysiological modalities and excellent skin biocompatibility, providing a new approach for constructing a health monitoring and smart medical system.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Liu et al. (2026) studied this question.

synapsesocial.com/papers/69e07e582f7e8953b7cbf5a3https://doi.org/10.1039/d6nh00022c
Ask AI
Helpful
Bookmark
Share
View Full Paper