Regression Using 20 Samples to Harmonize a Point‐of‐Care Analyzer With a Commercial Laboratory Analyzer for Feline Plasma Biochemistry Testing

View Full Paper
RBRandolph M. BaralBFBente FlatlandSJSusan M. Jaensch

Key Points

  • To assess the effectiveness of harmonization between a point-of-care analyzer and a commercial laboratory analyzer for feline plasma biochemistry.
  • Conducted regression analysis using 20 samples
  • Evaluated analytes for harmonization effectiveness
  • Identified limitations based on patient result ranges and relationships
  • Most analytes were successfully harmonized through this method
  • Harmonization failed for some due to inadequate concentration ranges
  • Non-linear relationships and absence of clinical performance goals contributed to harmonization challenges

Abstract

Most analytes could be harmonized by this approach. Reasons for harmonization failure were inadequate concentration range of patient results (AST, lipase, total bilirubin, GGT), potential non-linear relationship (lipase), and no clinical performance goal (globulins). Additional studies using samples with clinically important values may overcome these limitations, and similar studies of other species and other types of point-of-care analyzers would be of interest.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Baral et al. (2026) studied this question.

synapsesocial.com/papers/69e31f9e40886becb653ed1bhttps://doi.org/10.1111/vcp.70108