Coronary angioscopy allowed visualization of the coronary lumen and atheromatous plaque in 3 of 5 patients, though technical limitations restricted passage in the remaining 2.
Observational
Is it feasible to perform coronary angioscopy during cardiac catheterization to visualize the coronary lumen?
5 patients with obstructed right coronary artery after routine cardiac catheterization by the brachial approach
Introduction of a 5F Olympus Ultrathin fiberscope into the obstructed right coronary artery using a modified 8.3F USCI woven Dacron angioplasty guiding catheter
Feasibility of introducing the fiberscope and visualization of the coronary lumensurrogate
Coronary angioscopy is feasible for visualizing proximal coronary lesions but is limited by technical issues such as scope flexibility, lack of angulation, and image distortion.
Coronary angioscopy should permit direct inspection of the luminal cross section and identification of disease. The feasibility of introducing a 5F Olympus Ultrathin fiberscope into the obstructed right coronary artery in five patients after routine cardiac catheterization by the brachial approach was therefore tested. An 8.3F USCI woven Dacron angioplasty guiding catheter was modified to enlarge its lumen. After engaging the right coronary ostium with the catheter, an attempt was made to pass the angioscope coaxially to the tip of the catheter. Visualization of the coronary lumen was then achieved in three patients by manually injecting 5 to 10 cc of normal saline solution through the guiding catheter at 2 to 3 cc/s. White atheromatous plaque could be seen near the site of obstruction in each case. In two patients, a lack of sufficient flexibility in the distal 2 cm of the angioscope prevented passage to the catheter tip. Preliminary experience with a videoendoscopic system suggests that this monitoring technique is essential for the adequate performance of angioscopy and for recording dynamic changes during blood displacement. Geometric distortion of the image and nonlinearities in magnification and light reflex with a decreasing lens to object distance make quantitative evaluation of the lumen difficult. Lack of an angulation system further contributes to this problem and, more importantly, restricts passage of the angioscope to the proximal 1 to 2 cm segments of coronary arteries. Although coronary angioscopy may have research and clinical applications in the future, these technical problems should first be addressed.
Building similarity graph...
Analyzing shared references across papers
Loading...
J. Richard Spears
Artur M. Spokojny
H. John Marais
Journal of the American College of Cardiology
Beth Israel Deaconess Hospital
Building similarity graph...
Analyzing shared references across papers
Loading...
Spears et al. (Mon,) conducted a observational in Obstructed right coronary artery (n=5). Coronary angioscopy (5F Olympus Ultrathin fiberscope) was evaluated on Visualization of the coronary lumen. Coronary angioscopy allowed visualization of the coronary lumen and atheromatous plaque in 3 of 5 patients, though technical limitations restricted passage in the remaining 2.
www.synapsesocial.com/papers/69ee2ce19de2ebe493710045 — DOI: https://doi.org/10.1016/s0735-1097(85)80258-8