Hemodynamic support with the Impella 2.5 device during high-risk PCI was safe and feasible, with 30-day mortality of 5.5%, myocardial infarction of 0%, and stroke of 0.7%.
Observational
Yes
High-risk percutaneous coronary intervention (n=144)
Impella 2.5
Incidence of 30-day adverse events and successful device function
OBJECTIVES: This retrospective multicenter registry evaluated the safety and feasibility of left ventricular (LV) support with the Impella 2.5 (Abiomed Europe GmbH, Aachen, Germany) during high-risk percutaneous coronary intervention (PCI). BACKGROUND: Patients with complex or high-risk coronary lesions, such as last remaining vessel or left main lesions, are increasingly being treated with PCI. Because periprocedural hemodynamic compromise and complications might occur rapidly, many of these high-risk procedures are being performed with mechanical cardiac assistance, particularly in patients with poor LV function. The Impella 2.5, a percutaneous implantable LV assist device, might be a superior alternative to the traditionally used intra-aortic balloon pump. METHODS: The Europella registry included 144 consecutive patients who underwent a high-risk PCI. Safety and feasibility end points included incidence of 30-day adverse events and successful device function. RESULTS: Patients were older (62% >70 years of age), 54% had an LV ejection fraction < or = 30%, and the prevalence of comorbid conditions was high. Mean European System for Cardiac Operative Risk Evaluation score was 8.2 (SD 3.4), and 43% of the patients were refused for coronary artery bypass grafting. A PCI was considered high-risk due to left main disease, last remaining vessel disease, multivessel coronary artery disease, and low LV function in 53%, 17%, 81%, and 35% of the cases, respectively. Mortality at 30 days was 5.5%. Rates of myocardial infarction, stroke, bleeding requiring transfusion/surgery, and vascular complications at 30 days were 0%, 0.7%, 6.2%, and 4.0%, respectively. CONCLUSIONS: This large multicenter registry supports the safety, feasibility, and potential usefulness of hemodynamic support with Impella 2.5 in high-risk PCI.
Building similarity graph...
Analyzing shared references across papers
Loading...
Krischan D. Sjauw
Thomas Konorza
Raimund Erbel
Journal of the American College of Cardiology
University of Amsterdam
Amsterdam UMC Location University of Amsterdam
Copenhagen University Hospital
Building similarity graph...
Analyzing shared references across papers
Loading...
Sjauw et al. (Tue,) conducted a observational in High-risk percutaneous coronary intervention (n=144). Impella 2.5 was evaluated on Incidence of 30-day adverse events and successful device function. Hemodynamic support with the Impella 2.5 device during high-risk PCI was safe and feasible, with 30-day mortality of 5.5%, myocardial infarction of 0%, and stroke of 0.7%.
www.synapsesocial.com/papers/69ef7e653f5558efda2ad85a — DOI: https://doi.org/10.1016/j.jacc.2009.09.018