Impact of dose allocation between high‐ and low‐kV scans on virtual monoenergetic image quality in developmental dual‐energy cone‐beam CT for image‐guided radiotherapy

View Full Paper
AKAndrew KeelerLLLuke LaymanHNHa Nguyen

Key Points

  • This study evaluates how dose allocation between low- and high-energy scans affects image quality in dual-energy cone-beam CT for radiotherapy.
  • Conducted seven DE-CBCT scans with varying low/high energy dose allocations (10%/90% to 70%/30%).
  • Maintained a constant total imaging dose during the scans.
  • Reconstructed virtual monoenergetic images at varying keV levels for evaluation.
  • Protocols with dose allocations between 20%/80% and 70%/30% improved CNR by 18%-32% at 60 keV and reduced nonuniformity artifacts.
  • Specific dose allocation had a statistically insignificant impact on VMI quality between 20%/80% and 50%/50% allocations.
  • 70%/30% allocation significantly reduced HU accuracy compared to others, while 10%/90%, 60%/40%, and 70%/30% allocations showed reduced rCNR at specific energies.

Abstract

Abstract Background Dual‐energy cone‐beam CT (DE‐CBCT) offers potential as an advanced imaging technique for image‐guided radiation therapy (IGRT) through the production of virtual monoenergetic images (VMI). CBCT‐VMI can show enhanced image quality, with improved CNR and reduced nonuniformity artifacts, compared to polychromatic images. However, optimized DE‐CBCT protocols are required for clinical application. In particular, the impact of the dose allocation between the two acquisition energies on the resulting VMI image quality remains to be assessed. Purpose This study evaluates the impact of different dose allocation strategies between the low‐ and high‐energy scans on the image quality of DE‐CBCT derived VMI. Methods A series of seven DE‐CBCT scans were performed using various dose allocations ranging from 10% Low‐energy/90% High‐energy (LE/HE) to 70%/30% LE/HE in 10% intervals while maintaining a constant total imaging dose. To assess the impact of the dose allocation strategies, VMI were reconstructed at 40, 60, 80, and 100 keV and evaluated for Hounsfield Unit (HU) accuracy, relative contrast‐to‐noise ratio (rCNR) enhancement, and HU uniformity compared to a dose‐matched clinical CBCT protocol. Results DE‐CBCT protocols between 20%/80% and 70%/30% LE/HE showed significant improvements over the clinical protocol, with CNR values improving by an average of 18%–32% at 60 keV and non‐uniformity artifacts reduced to negligible levels without loss of HU accuracy. Statistical comparison between these DE‐CBCT protocols further showed that the specific dose allocation had statistically insignificant impact on the resulting VMI image quality between 20%/80% and 50%/50% LE/HE at any energy. Of the remaining DE protocols, 70%/30% showed a significant decrease in HU accuracy while 10%/90%, 60%/40%, and 70%/30% showed reduced rCNR enhancement at some energies. Conclusions Overall, a wide range of dose allocations are shown to produce comparably near‐optimal VMI quality. This finding supports protocol design flexibility for future clinical DE‐CBCT implementations.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Keeler et al. (2026) studied this question.

synapsesocial.com/papers/69f1545d879cb923c494476dhttps://doi.org/10.1002/acm2.70545
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Technical note: TIGRE‐DE for the creation of virtual monoenergetic images from dual‐energy cone‐beam CT2024 · 4 citations
  2. 2Multi‐energy blended CBCT spectral imaging and scatter‐decoupled material decomposition using a spectral modulator with flying focal spot (SMFFS)2024 · 6 citations
  3. 3Fast, automated optimization of virtual monoenergetic images with the dual‐energy image synthesizer for cone‐beam CT2025 · 3 citations
  4. 4Combined iterative reconstruction and image-domain decomposition for dual energy CT using total-variation regularization2014 · 72 citations
  5. 5Quantitative accuracy and dose efficiency of dual‐contrast imaging using dual‐energy CT: a phantom study2019 · 24 citations