April 30, 2026Journal of Neuroendovascular Therapy0 citationsOpen Access

Intra-Distal Access Catheter Deployment of the Surpass Evolve Device: A Novel Technique to Prevent Advancement of the Delivery Wire: A Case Report and Literature Review

MSMiku SugaiHNHiroaki NekiYKYoshinobu Kamio

Key Points

  • The aim is to present a novel technique to enhance safety during flow diverter stent deployment, particularly for distal aneurysms.
  • Described a deployment strategy using a distal access catheter
  • Mitigated wire advancement risk during flow diverter stent deployment
  • Reduced risk of vessel injury during FDS deployment
  • Enhanced control of delivery wire behavior

Abstract

Objective: Flow diverter stent (FDS) deployment in neuroendovascular therapy carries a risk of vessel injury related to unintended distal advancement of the delivery wire, particularly in cases with distal aneurysms. Safe control of wire behavior during deployment remains challenging. We describe a deployment strategy using a distal access catheter (DAC) to mitigate this risk.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sugai et al. (2026) studied this question.

synapsesocial.com/papers/69f2f0991e5f7920c6386bb5https://doi.org/10.5797/jnet.cr.2026-0010
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Endovascular Rescue of Previously Treated Carotid Aneurysm: Higher Radial Force 48-Wire Flow Diverter for Salvage of Collapsed 64-Wire Device With Braid Deformation: 2-Dimensional Operative Video2024 · 1 citations
  2. 2Abstract 513: Surpass Evolve Flow Diverter For Intracranial Aneurysms: Single‐Center Experience2025
  3. 3A rescue treatment to release the twist of a flow re-direction endoluminal device2024
  4. 4The Revolve Study: A Multicenter Registry Using the Flow Diverter Surpass Evolve® Device for the Endovascular Treatment of Intracranial Aneurysms2026
  5. 5Delayed Aneurysmal Rupture Following Flow Diverter Placement despite adjunctive coiling: Case report2026