May 4, 2026European Heart Journal93 citationsOpen Access

A trial of self-adhesive patch electrodes and hand-held paddle electrodes for external cardioversion of atrial fibrillation (MOBIPAPA)

View Full Paper
PKPaulus KirchhofGMGerold MönnigKWKristina Wasmer

Structured PICO

Do hand-held paddle electrodes improve cardioversion success compared to adhesive patch electrodes in patients with persistent atrial fibrillation?

P
Population
201 consecutive patients with persistent atrial fibrillation (147 male, mean age 63+/-1 years, duration of atrial fibrillation 6.3+/-1 months)
I
Intervention
Hand-held steel paddle electrodes for external cardioversion (anterior-posterior position)
C
Comparator
Adhesive patch electrodes for external cardioversion (anterior-posterior position)
O
Outcome
Cardioversion success

Hand-held paddle electrodes significantly increase the success rate of external cardioversion for persistent atrial fibrillation compared to adhesive patch electrodes.

Abstract

AIMS: External electrical cardioversion is the method of choice to terminate persistent atrial fibrillation. Whether the type of shock electrode affects cardioversion success is not known. We tested whether hand-held steel electrodes improve cardioversion outcome with monophasic or biphasic shocks when compared with adhesive patch electrodes. METHODS AND RESULTS: Two hundred and one consecutive patients with persistent atrial fibrillation (147 male, mean age 63+/-1 years, duration of atrial fibrillation 6.3+/-1 months) were randomly assigned to cardioversion using either a sinusoidal monophasic or a truncated exponential biphasic shock wave form. The first half of patients were cardioverted using adhesive patch electrodes, the second half using hand-held steel paddle electrodes, and all patients using an anterior-posterior electrode position. Paddle electrodes successfully cardioverted 100/104 patients (96%) and patch electrodes 85/97 patients (88%, P=0.04). This effect was comparable to that of biphasic shocks: biphasic shocks cardioverted 102/104 patients (98%) and monophasic shocks 83/97 patients (86%, P=0.001). A beneficial effect of paddle electrodes was observed for both shock wave forms. After cross-over from an ineffective monophasic to a biphasic shock, cardioversion was successful in 198/201 (98.5%) patients. Unsuccessful cardioversion after cross-over (3/201 patients) only occurred with patch electrodes (P=0.07). CONCLUSION: Hand-held paddle electrodes increase success of external cardioversion of atrial fibrillation in this trial. This increase is of similar magnitude as the increase in cardioversion success achieved with biphasic shocks. A combination of biphasic shocks, paddle electrodes, and an anterior-posterior electrode position renders outcome of external cardioversion almost always successful (104/104 patients in this trial).

Expert Takes1 quote

“The proportion of patients who cardioverted on the first shock was almost two times higher in the group after the first shock with the paddles (50% vs. 27%; P = .01)”

Jose A. Joglar, Cardiac electrophysiology program director, UT SouthwesternUniversity of Texas Southwestern Medical Centerauto_pipelineSupportiveView source
Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Kirchhof et al. (2005) studied this question.

synapsesocial.com/papers/69f8b8e1037bf4ee0479fb65https://doi.org/10.1093/eurheartj/ehi160