Key result
Jailing transvenous pacemaker leads during TTVR can trigger subacute complete heart block and lead malfunction.
Why the study?
The short- and long-term effects of lead entrapment during transcutaneous tricuspid valve replacement on lead function are not well understood due to limited studies with low enrollment.
Case Report (n=1)
No
Jailing preexisting transvenous leads during TTVR can cause unpredictable, life-threatening subacute lead compromise and complete heart block, highlighting the need for multidisciplinary planning and potential prophylactic pacing.
Answers are grounded in this paper and cite their sources.
Warrants caution with preexisting leads during TTVR; leaves open optimal periprocedural lead strategies.
Transcutaneous tricuspid valve replacement (TTVR) is one of the expanding procedures in the field of structural interventional cardiology to treat severe tricuspid regurgitation (TR) and frequently employed when edge-to-edge repair is not deemed feasible.1 Roughly one-third of patients referred for TTVR have pre-existing cardiac implantable electrical devices (CIEDs) which may or may not be causally linked to TR itself, the management of which includes transvenous lead extraction ahead of TTVR or jailing of the current leads.
No takes yet. Share an insight, caveat, or question.
Bogun et al. (2026) conducted a case report in Severe tricuspid regurgitation with preexisting cardiac implantable electrical device (n=1). Transcutaneous tricuspid valve replacement with jailing of transvenous lead was evaluated on Subacute complete heart block with concomitant lead malfunction. Jailing of a preexisting transvenous pacemaker lead during transcutaneous tricuspid valve replacement resulted in subacute complete heart block and life-threatening lead malfunction.
Synapse has enriched 10 related papers. Consider them for comparative context: