May 28, 2026The European Physical Journal Special Topics0 citationsOpen Access

Feasibility study of adapting national postal TLD dosimetry audits for tomotherapy units: the Polish experience

WŚWioletta Ślusarczyk‐KacprzykIGIwona GrabskaMSMarcin Szymański

Key Points

  • This study aims to explore the feasibility of adapting dosimetry audits for tomotherapy units to enhance treatment accuracy.
  • Conducted feasibility study in 2022 and 2025 using fifteen thermoluminescent detectors on the Accuray RADIXACT device.
  • Modified audit methodology to accommodate helical therapy's unique requirements and measurement conditions.
  • Evaluated percentage deviation of measurements in accordance with established tolerance ranges.
  • Percentage deviation for individual detectors ranged from -1.2% to 0.8%, mean -0.2%.
  • Mean deviation for measurement sets was between -0.6% and 0.3%, overall average -0.1%.
  • All results were within the tolerance range of ±3.5%, supporting the use of helical techniques clinically.

Abstract

Abstract The continuous development of radiotherapy equipment and techniques requires existing dosimetry audits to adapt to these changes. One advanced form of radiotherapy is helical therapy, commonly referred to as tomotherapy. During helical therapy, a linear accelerator rotates around the patient, delivering radiation similar to a spiral CT scan. Simultaneously moving the treatment couch during irradiation allows large areas to be irradiated without combining fields. This is important for treating gynaecological, soft tissue and bone cancers, as well as for using special techniques such as total body irradiation (TBI), half-body irradiation (HBI) and total marrow irradiation (TMI). In 2022 and 2025, feasibility study of adapting TLD dosimetry audit of tomotherapy was conducted on the Accuray RADIXACT device at the Maria Sklodowska-Curie National Research Institute of Oncology in Warsaw. The study, performed using fifteen thermoluminescent (TL) detectors, yielded highly satisfactory results. The percentage deviation for individual detectors ( delta i ) ranged from − 1.2% to 0.8%, with a mean value of − 0.2% (SD = 0.6%). For the measurement sets, the mean deviation ( delta mean ) ranged from − 0.6% to 0.3%, with an overall average of − 0.1% (SD = 0.4%). All results were well within the required tolerance range of ± 3.5%. The positive outcome of this study performed using thermoluminescent (TL) detectors was one of the factors contributing to the approval of the helical technique for clinical use at our centre. The audit methodology for conventional accelerators has been modified because it was impossible to create the required conventional reference conditions (the source-axis distance of 100 cm and a field size of 10 cm × 10 cm) when using helical techniques. The study demonstrates the feasibility of adapting existing postal dosimetry protocols for unconventional radiotherapy devices.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Ślusarczyk‐Kacprzyk et al. (2026) studied this question.

synapsesocial.com/papers/6a17ddab3fad632b0f9da59dhttps://doi.org/10.1140/epjs/s11734-026-02394-9
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Targeted Total Marrow Irradiation Using Three-Dimensional Image-Guided Tomographic Intensity-Modulated Radiation Therapy: An Alternative to Standard Total Body Irradiation2006 · 222 citations
  2. 2Clinical feasibility of TBI with helical tomotherapy2010 · 87 citations
  3. 3Lithium Fluoride Thermoluminescence Dosimetry1964 · 73 citations
  4. 4Half-Body Irradiation With Tomotherapy for Pain Palliation in Metastatic Breast Cancer2013 · 16 citations
  5. 5[P162] Postal TLD dosimetry audits in radiotherapy in poland – Results of 20172018 · 2 citations