Evaluation of Four Automated Nucleic Acid Testing Systems for the Ultrasensitive Quantification of Hepatitis B Virus

View Full Paper
ZHZhongqiang HuangLYLiufan YinZLZhenhua Liu

Key Points

  • The aim is to evaluate and compare four automated testing systems for quantifying hepatitis B virus to promote standardization.
  • Assessed linearity with 10-fold serial dilutions of HBV concentrate.
  • Evaluated precision, accuracy, and sensitivity using WHO standards.
  • Analyzed clinical performance using 157 clinical specimens.
  • Roche showed linearity from 1.0×10^1 to 2.0×10^8 IU/mL; others stable only above 100 IU/mL.
  • Roche and Tianlong achieved the highest sensitivity of 2.5 IU/mL; Daan and Sansure achieved 5 IU/mL.
  • All systems had excellent linear correlation with Roche (r > 0.97), with small mean differences in IU/mL.

Abstract

Objectives: To comprehensively evaluate and compare four commonly used automated nucleic acid testing systems (Roche, Tianlong, Daan, and Sansure) for the quantification of hepatitis B virus, and to promote cross-platform standardization. Methods: Linearity was assessed using 10-fold serial dilutions of a high-concentration HBV concentrate. Precision, accuracy, and analytical sensitivity were evaluated using WHO international standards. Clinical performance was assessed using 157 clinical specimens. Results: Roche exhibited good linearity over 1.0×10 1 -2.0×10 8 IU/mL, while Tianlong, Daan, and Sansure showed stable linear detection capability only at concentrations above 100 IU/mL. All four systems exhibited acceptable precision and accuracy, among which Roche performed the best in low-concentration samples. Roche and Tianlong achieved the highest analytical sensitivity of 2.5 IU/mL, while Daan and Sansure reached 5 IU/mL. Using Roche as the reference, the clinical performance of the other three systems was evaluated. These systems exhibited excellent linear correlation (r > 0.97) with Roche, with mean differences of 0.383, 0.228, and -0.111 log 10 IU/mL, respectively. Conclusions The four systems differed in performance, with Roche demonstrating the best overall performance, especially at low viral loads. The other three systems showed satisfactory correlation and concordance with Roche and are suitable for resource-limited settings.