June 3, 2026The Kitakanto Medical Journal0 citationsOpen Access

Miscorrection Risk in Motion Correction for Low-Count Nuclear Images Simulating Pediatric Renal Imaging:

HHHiromitsu HoshinoSSSuzue Sato

Key Points

  • This study aims to quantitatively evaluate the risk of miscorrection during motion correction in pediatric renal imaging under low-count conditions.
  • Utilized a stationary NEMA IEC body phantom to simulate pediatric renal imaging conditions.
  • Assessed motion correction software's impact on image quality under low-count scenarios.
  • Identified significant miscorrection risks leading to image degradation in low-count nuclear images.
  • Quantitative analysis demonstrated that statistical noise was often misinterpreted as motion, affecting image clarity.

Abstract

背景・目的:核医学画像における体動補正ソフトウェアは広く臨床応用されているが,低カウント条件では統計ノイズを体動と誤認識し,かえって画質を劣化させる誤補正が生じる可能性がある.特に小児99mTc-DMSA腎静態シンチグラフィでは投与量が制限されるため低カウント条件となりやすく,この問題は臨床的に重要である.本研究では,小児腎静態イメージングを模した条件下で,静止状態のNEMA IEC bodyファントムを用いて体動補正の誤補正リスクを定量的に評価することを目的とした.