Answers are grounded in this paper and cite their sources.
This study presents the results of a multicentre external dosimetry audit conducted as part of the PROTECT trial, involving nine European radiotherapy centres delivering proton and photon treatments. A dosimetry equipment vendor performed beam output audits at participating sites, verifying the absorbed dose-to-water measurements. Measurements used reference conditions based on the IAEA TRS-398, with ionization chambers traceable to primary standards. The audit compared local centre measurements to independent assessments, finding all within the acceptable 3% agreement threshold. The study is part of the successful implementation of a comprehensive, multicentre QA programme in the PROTECT trial.
No takes yet. Share an insight, caveat, or question.
Ehmsen et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: