An analysis of the flow data demonstrated statistically significant differences between combinations of HBV-AGV settings adding up to the same theoretical opening pressure. These findings suggest that in practice, the nominal OPs of Hakim and M.blueⓇ valves may not be additive. Clinicians should be aware that drainage patterns can change when adjusting OPs of HakimⓇ and M.blueⓇ valves under the assumption that HBV-AGV settings with the same summed theoretical OP behave alike.